ANALYZER , BLOOD GAS
Definition:
AN INSTRUMENT DESIGNED FOR DETERMINATION OF THE CARBON DIOXIDE COMBINING POWER, OXYGEN, CARBON MONOXIDE AND NITROGEN OF WHOLE BLOOD OR BLOOD PLASMA. MAY ALSO BE USED FOR MICRO DETERMINATIONS OF GAS OTHER THAN BLOOD GASES, DEVELOPED FROM SOLUTIONS, SUCH AS ORGANIC NITROGEN AND CARBON, UREA, AMINO NITROGEN, IODATES, SULPHATES, SUGAR, CALCIUM LACTIC ACID, POTASSIUM, AND ANALYSIS OF GAS MIXTURES.
CAGE Information
Code | Company |
---|---|
0CNN1 | MALLINCKRODT SENSOR SYSTEMS INC |
Federal Supply Class
Title
CHEMICAL ANALYSIS INSTRUMENTS
Inclusions:
INCLUDES ACIDITY (PH) METERS; GAS ANALYZERS; ALAKALINITY MEASURING INSTRUMENTS; COLORIMETERS.
Exclusions:
EXCLUDES GAS DETECTING EQUIPMENT; MANUAL GAS ALARMS; CHEMICALS; IN VITRO DIAGNOSTIC SUBSTANCES AND REAGENTS.
Characteristics (Decoded)
MRC | Requirements Statement | Clear Text Reply |
---|---|---|
CXCY | PART NAME ASSIGNED BY CONTROLLING AGENCY | BLOOD GAS ANALYZER 120/230V 50/60 HZ AC: |
CYFY | ANALYSIS TEST/MEASUREMENT RANGE | XPO2 0 TO 600 MMHG; PH 6.8 TO 7.8; PCO2 8 TO 99 MMHG; K 0.1 TO 9.9 MMOL/L; NA 100 TO 200 MMOL/L; CA 0.25 TO 4.25 MMOL/L; HCT 10 TO 75 PCT; HC03 3.0 60 MMOL/L TCO2 3.0 TO 60 MMOL/L; BASE EXCESS -30 TO +30 MMOL/L; O2 SAT 30 TO 90 PCT. |
CYFV | METHOD OF ANALYSIS | POLAROGRAPHIC AND POTENTIOMETRIC/ION SELECTIVE ELECTRODE |
EXME | EXTERIOR CONTAINER CUBIC MEASURE | 1.00 CUBIC FEET |
ABRN | HEIGHT | 8.400 INCHES |
WDTH | WIDTH | 9.100 INCHES |
AQXY | TEST TYPE FOR WHICH DESIGNED | BLOOD GASES AND ELECTROLYTES,HC03,HCT,BASE EXCESS AND 02 SAT |
ECWT | EXTERIOR CONTAINER WEIGHT | 27.00 POUNDS |
ELEC | VOLTAGE IN VOLTS | 120.0 AND 230.0 |
CBBL | FEATURES PROVIDED | INTEGRAL PRINTER AND DIGITAL DISPLAY SCREEN AND SELF-CALIBRATION |
CYFZ | TEST SAMPLE/SPECIMEN SIZE | 0.5 ML |
CYHM | TEST SAMPLE/SPECIMEN TYPE | WHOLE BLOOD |
ACDC | CURRENT TYPE | AC |
FEAT | SPECIAL FEATURES | UTILIZES A DISPOSABLE CARTRIDGE/TEST PACK THAT CONTAINS ALL ELECTRODES, CALIBRATING SOLUTIONS AND GASES PLUS A WASTE BAG FOR PROCESSED BLOOD, CALIBRATING AND QC SOLUTIONS. THE CARTRIDGE/TEST PACK SHALL ELIMINATE THE NEED FOR GAS TANKS EXTERNAL SOLUTIONS, AND ELECTRODE MAINTENANCE. SHALL BE PRE-PROGRAMED TO AUTOMATICALLY PERFORM TWO-POINT CALIBRATIONS EVERY HOUR AND ONE-POINT CALIBRATIONS EVERY 90 SECONDS. CAPABLE OF AUTOMATIC ASPIRATION OF THE SAMPLE AND TO PROCESS, DISPLAY, AND PRINT THE RESULTS IN LESS THAN 2 MINUTES. SHALL BE SUITABLE FOR THE FIELD ENVIRONMENT. |
ACZB | FREQUENCY RATING | 50.0 HERTZ NOMINAL OR 60.0 HERTZ NOMINAL |
DPTH | DEPTH | 18.300 INCHES |